Light Source |
|
|
Lamp |
150 W Xe arc lamp |
|
Lamp Housing |
Air-cooling, Shutter, Iris, Optical filter holder |
|
Stopped-flow Mixer with an Automatic Cock |
|
|
Mixing Dead Time |
3 msec or less (10 mm cell), 0.9 msec (10 mm microcell) |
|
Mixer |
Double 2 JET |
|
Mode |
Piston-drive, Front-stop |
|
Cells |
10 mm cell for absorption measurement. 2 mm cell for fluorescence measurement (option) |
|
Mixing Ratio |
1:1 (up to 1:9 available by option) |
|
Sample Amount |
75μL ~ 250 μL/shot |
|
Temperature Range |
5°C ~ 60°C |
|
Multi-Channel Detection |
|
|
Spectrometer |
Polychromator PK120 |
|
Dispersion Element |
Aberration-corrected Concave Grating |
|
Brightness |
F/2.9 |
|
Linear Dispersion |
22 nm/mm |
|
Slit Width |
0.05 mm, 0.1 mm, 0.2 mm |
|
Second Diffraction Cut-off |
Dedicated filter is incorporated |
|
Detector |
Photodiode array 512ch |
|
Maximum Scan Speed |
1 msec ~ 10 sec/scan |
|
Number of Scan |
Up to 8000 |
|
Spectral Width |
560 nm* Specify central wavelength at the time of order. |
|
Monochromator Detection |
|
|
Spectrometer |
Monochromator MF200 |
|
Dispersion Element |
Plane Blazed Holographic Grating |
|
Brightness |
F/3.5 |
|
Linear Dispersion |
8.5 nm/mm |
|
Slit Width |
0.1 mm, 0.2 mm, 0.5 mm |
|
Second Diffraction Cut-off |
None |
|
Detector |
Hamamatsu PMT R2949 |
|
Filter Time Constant |
10 μs ~ 600 ms |
|
Sampling Time |
10 μs ~ 600 ms |
|
Optical Guides |
|
|
Material |
UV-transmittable quartz |
|
Length |
1 m |
|
Data Processor |
|
|
OS |
Windows 7/XP/2000 |
|
Interface |
PCI bus |
|
Software |
Automatic washing, non-linear least square fitting, data storing by text file |
| 点击观看pdf资料 | ![]() |